Mass spectrometry: how to prove the vial contains the right peptide

If HPLC purity answers the question "how pure is it?", mass spectrometry (MS) answers the question "what is it?". It measures the mass of molecules. Since the mass of a peptide can be calculated exactly from its sequence, a match between the measured and the calculated mass is the basic proof of identity.
How a mass spectrometer works
The instrument does not measure mass directly. It first converts molecules into ions, that is, gives them a charge, and then measures the ratio of mass to charge, written as m/z. Two ionisation methods are most commonly used for peptides:
- Electrospray (ESI): the solution is sprayed into fine droplets under high voltage. In the process, peptides often pick up several protons, so the same molecule appears as several signals, with one, two or three charges.
- MALDI-TOF: the peptide is mixed with a matrix and exposed to a laser pulse. Here the peptide is usually seen with a single charge.
Electrospray is often coupled directly to an HPLC system. This combination is called LC-MS, and it gives every peak in the chromatogram its mass as well.
Example: the tripeptide GHK
The tripeptide GHK (glycine, histidine, lysine) has the molecular formula C14H24N6O4. Its average molar mass is 340.38 g/mol, and its monoisotopic mass, calculated only from the most abundant isotope of each element, is 340.19.
In an ESI spectrum we therefore expect:
- a signal [M+H]+ at m/z 341.19, the molecule with one added proton,
- a weaker signal [M+2H]2+ at m/z 171.10, with two protons, so the m/z value is halved.
High-resolution instruments also show the isotope pattern, a series of smaller peaks next to the main one. The spacing between them reveals the charge: it is 1 for a singly charged ion and 0.5 for a doubly charged one.
Signals that confuse but are normal
Besides the expected ions, the spectrum often also shows adducts: the same molecule with sodium, [M+Na]+, which is 22 units heavier than [M+H]+, or with potassium, [M+K]+, heavier by 38. Sodium is everywhere, so these signals are not a sign of a bad peptide. What matters is that the laboratory states which ion it took as confirmation.
What mass spectrometry cannot see
- Same composition, different order. The peptides GHK and KHG have exactly the same mass. To confirm the order, the molecule is broken into fragments inside the instrument (tandem mass spectrometry, MS/MS), and the sequence is read from the masses of those fragments.
- Isomers. Leucine and isoleucine have the same mass, as do the L- and D-forms of the same amino acid. Ordinary MS cannot tell them apart.
- Quantity. Different molecules do not ionise equally easily, so the height of a signal does not tell you how much of what is present. Purity requires HPLC.
When MS is particularly useful
The combination LC-MS provides the most information. If there is an impurity next to the main peak, its mass often reveals what it is: a chain missing one amino acid is lighter by exactly the mass of that one link, and an oxidised methionine is heavier by 16 units. This also shows where the impurity comes from, synthesis or degradation.
What a good MS report looks like
- The theoretical and the measured mass are stated, as well as the type of ion.
- The difference between them is within the accuracy of the instrument.
- The spectrum is attached or can be obtained on request.
- The batch number matches the vial.