How to read a chromatogram: from the baseline to the hidden peak

A certificate of analysis (COA) often comes with an image: a line that runs mostly flat and then, at one point, rises sharply into a tall peak. This is the chromatogram, the raw record of the HPLC analysis that produces the purity figure. Once you know what to look for, you can judge for yourself how convincing that figure is.

What the axes show

The horizontal axis is time, usually in minutes, counted from the moment the sample is injected onto the column. The vertical axis is the detector signal. For peptides this is most often a UV detector, so the signal is expressed in absorbance units, often as mAU. While only the mobile phase is leaving the column, the signal stays low and steady. When a substance that absorbs light comes through, the signal rises and falls again, and that is how a peak forms.

Retention time and the baseline

Retention time is the moment at which the top of the peak appears. Under the same conditions, the same substance elutes at roughly the same time, so peaks can be recognised by their retention time and compared with a reference standard. Retention time alone does not prove identity, however, because two different substances can elute together. That is what mass spectrometry is for.

The baseline is the signal recorded when nothing the detector can see is leaving the column. In a gradient analysis it may rise slightly as the composition of the mobile phase changes, and that is expected. Noise, waviness and sudden jumps are a problem: the baseline determines where a peak begins and ends, and therefore its area.

Area, percentages and small peaks

For each peak, the software calculates the area under the curve. Purity is the main peak's share of the combined area of all peaks, the so-called area percent. If the main peak makes up 98.5 % of the total area, the result reads 98.5 %. The calculation assumes that all substances absorb light roughly equally, which is a reasonable but not a perfect assumption.

Small peaks before and after the main one are most often impurities. In synthetic peptides these are usually related molecules: chains missing one amino acid, truncated chains, or chains still carrying a protecting group left over from synthesis. Small peaks are normal. What matters is that they are visible and included in the calculation.

Peak shape

Sharp and symmetrical

A good main peak is narrow and symmetrical: it rises quickly, falls quickly, and its two sides are roughly the same. A narrow peak means the column is working efficiently, which makes separation from neighbouring substances easier.

A tailing peak

When the back of a peak drags out and descends slowly towards the baseline, the peak is said to tail. The cause usually lies in the method: interaction with the column, a worn column or too much sample. The European Pharmacopoeia and the USP describe peak symmetry with a dedicated factor, which equals one for a perfectly symmetrical peak. Tailing does not necessarily mean the sample is poor, but it makes integration harder, and a small impurity can get lost in the tail.

The shoulder and what it can hide

A shoulder is a bulge on the flank of the main peak, as if a second, smaller peak were leaning against it. Most often that is exactly what it is: another substance eluting at almost the same time as the main one. If the two substances are not fully separated, the software often integrates them as a single peak. The impurity's area is then attributed to the main peak, and purity looks higher than it really is. Molecules that differ from the target by a single amino acid behave very similarly in the column, so they are precisely the ones most likely to stay hidden in a shoulder. If a shoulder is visible but the peak table lists only one peak, that is a reason to ask questions.

A short checklist for reading a chromatogram

  • The batch number on the certificate and the chromatogram should match the number on the label of the vial you are holding.
  • The axes should be labelled: time on the horizontal axis, detector signal on the vertical one.
  • The baseline should be flat or gently sloping, without sudden jumps.
  • The main peak should be narrow and roughly symmetrical, with no visible shoulder.
  • Small peaks should be visible and listed in the peak table.
  • The area percent of the main peak in the table should match the purity stated on the certificate.
  • The method, column and wavelength should be stated alongside the chromatogram.